The device was evaluated by olympus.The evaluation was unable to confirm the reported event.The investigation uncovered a worn-out scope socket slider switch which caused intermittent use of high intensity mode.The olympus lamp was over 500 hours which was below specification.Additionally, the top metal cover was also found to be exposed.The faulty parts were replaced to meet olympus specifications.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
|