• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON Back to Search Results
Model Number 97810
Device Problems Display or Visual Feedback Problem (1184); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/30/2020
Event Type  malfunction  
Manufacturer Narrative
Analysis of the ins 97810 (s/n: (b)(4)) determined that the service life of the implantable neurostimulator (ins) was started prematurely.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer representative (rep) regarding a patient with an implanted neurostimulator (ins) for urinary dysfunction/sacral nerve stim and gastrointestinal/pelvic floor therapy.The rep reported that an account received an ins on (b)(6) 2020 but when the caller interrogated the ins it was only 10% charged.The caller charged the ins to 40% and then they received a por message.The caller tapped ok to advance into the session and there was already patient information entered and it had patient's dob as (b)(6) 2020.The caller speculated the ins may have not been used for a case and sent back, then shipped out again to another customer.The caller looked up the patient but didn't see them registered and assumed that (b)(6) 2020 may have been the date of the original case in which the ins wasn't used.Troubleshooting was not required.Technical services (ts) reviewed that the ins was likely okay to use given that the ins charged fine and devices can incur pors, but given the suspicious circumstances, the caller should return ins for analysis.Ts emailed caller with case number to use for customer service purposes.Ts changed case priority to critical in case of an out of box issue.There was no patient involvement in this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key11922754
MDR Text Key267101495
Report Number3004209178-2021-08704
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00763000203849
UDI-Public00763000203849
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/14/2022
Device Model Number97810
Device Catalogue Number97810
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2021
Initial Date Manufacturer Received 05/07/2021
Initial Date FDA Received06/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-