Fill volume: 500 ml, flow rate: unknown, procedure: knee replacement, cathplace: adductor canal block, infusion start time: (b)(6) 2021, time not provided.Infusion stop time:(b)(6) 2021, 0800.It was reported that the device delivered local anesthetic at a fast flow rate.The pump was set up for a continuous, surgeon-placed adductor canal block and when checked on the morning of (b)(6) 2021 was found to be almost depleted.The pump was filled to 500ml, however the red tab was not removed.It was identified that a new nurse set up the pump the day prior.The pump was clamped when the fast flow was identified.No patient harm was reported.Additional information received 13-may-2021 indicated the site has been using on-q since 2013, and the on demand model since 2018.
|