Model Number 3100A |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/12/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
|
Event Description
|
The customer reported that the 3100 a ventilator experienced its readings are fluctuating.The customer confirmed that there was no patient harm associated with the reported event.As an intervention, ventilator was removed and replaced with another.
|
|
Manufacturer Narrative
|
Result of investigation: a vyaire field service representative (fsr) went onsite and evaluated the ventilator.Panel meter ass'y (mean pressure monitor) was replaced.Passed all tests.Vent is operational and meets oem specs.
|
|