Model Number ACL TIGHTROPE WITH FIBERTAG
Device Problem Output Problem (3005)
Patient Problem Insufficient Information (4580)
Event Date 05/20/2021
Event Type  Injury  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
It was reported, that during an acl reconstruction with quad tendon harvest, after insertion on the ar-1588rtt, the surgeon notice the button on the tight rope was not loaded properly, the shuttle suture was wrapped around the button.The surgeon removed the tightrope repositioning the suture and the case was completed without further issue.The case was extended by 30 minutes adding additional anesthesia.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACL TIGHTROPE WITH FIBERTAG
Type of Device
PIN, FIXATION, SMOOTH
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key11946704
MDR Text Key254800026
Report Number1220246-2021-03210
Device Sequence Number1
Product Code HTY
UDI-Device Identifier00888867308459
UDI-Public00888867308459
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112990
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 06/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberACL TIGHTROPE WITH FIBERTAG
Device Catalogue NumberAR-1588RTT
Device Lot Number10928254
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/20/2021
Initial Date FDA Received06/07/2021
Date Device Manufactured05/06/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-