The codman ventricular microsensor was returned for evaluation: device history record (dhr) - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - the catheter was evaluated, and the following observations were noted: received catheter in a drainage tube.The sensor lead discolored.The sealant around sensor lifting.The icp express read ¿no transducer detected¿.There was no testing possible.Based on the evaluation, the complaint is confirmed.It is suspected that fluid entered at the lifting of the sealant around the sensor and caused corrosion.Per the complaint background, could not detect, no readings.The complaint was confirmed in the failure analysis.It is suspected that fluid entered at the lifting of the sealant around the sensor and caused corrosion.It is suspected that lifting of sealant was caused from inserting into drainage tube.As a result, no testing was possible.The root cause is ¿use related¿.Currently the complaint issue does not represent an adverse trend, however the issue will continue to be monitored and trended through the complaint evaluation process.
|