MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION VALVE, O.P.S BULK, N-S; OVERPRESSURE SAFETY VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION VALVE, O.P.S BULK, N-S; OVERPRESSURE SAFETY VALVE Back to Search Results
Model Number LN130B
Device Problems Leak/Splash (1354); Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2021
Event Type  malfunction  
Manufacturer Narrative
The returned sample was inspected and found the negative umbrella not seated properly in the housing causing the ops valve to not suction during the event as air was leaking past the negative umbrella rather than suctioning blood.The retention sample was visually inspected with no anomalies noted on the device.All ops valves are subjected to a 100% leak test that undergoes 5 separate programs to test that each component in the valve is functioning properly.These units are also 100% visually inspected both during the leak testing step and during packaging.(b)(4).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Event Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, the valve would not allow blood to pass through.Per facility, they tried clamping on either or both sides of the valve the blood was spurting through the umbrella valve, he was able to make it work properly after much manipulation.It was located on the aortic root vent line.No known consequence or impact to patient.There was 10-20cc of blood loss.The product was not changed out.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE, O.P.S BULK, N-S
Type of Device
OVERPRESSURE SAFETY VALVE
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key11956744
MDR Text Key264340960
Report Number1124841-2021-00133
Device Sequence Number1
Product Code DTL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K820297
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Model NumberLN130B
Device Catalogue NumberN/A
Device Lot NumberZA04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/27/2021
Initial Date FDA Received06/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-