• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2
Device Problem Material Frayed (1262)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device referenced in this report has been evaluated by olympus.Preliminary findings are reported.The investigation is ongoing.The definitive cause of the user¿s experience cannot be determined at this time.Physical evaluation of the complaint device reveals: confirmed customer complaint of bending rubber torn, ruptured, broken with metal sticking out.Leaking at channel and distal sheath of scope.The endoscope fully disassembled and reassembled with new insertion tube unit.Additional repairs to restore to original factory specifications.Endoscope leak tested to ensure water tight integrity.Endoscope restored to original factory specifications.This report will be updated upon completion of the investigation or upon receipt of additional relevant information.
 
Event Description
User reported uretero-reno videoscope failed leak test, with air/water leakages or fluid invasion.The bending rubber is broken/torn/ruptured and the metal is sticking out.No patient harm or injury occurred due to this malfunction.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The following sections were updated.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the phenomenon likely occurred because the user manipulated the device so that the bending section receives excessive stress.The user can properly detect the event by handling the device in accordance with the following statements, identified in the instructions for use: "precautions : perform a leakage test on the endoscope after each precleaning procedure.Do not use the endoscope if a leak is detected.Use of an endoscope with a leak may cause a sudden loss of the endoscopic image, damage to the bending mechanism, or other malfunctions.Use of a leaking endoscope may also pose an infection control risk." "inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities." the user can reduce / prevent occurrence of the event by handling the device in accordance with the following statements, identified in the instructions for use: "do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist." "do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11971104
MDR Text Key280677839
Report Number8010047-2021-07308
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170343582
UDI-Public04953170343582
Combination Product (y/n)N
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 07/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/12/2021
Initial Date FDA Received06/09/2021
Supplement Dates Manufacturer Received07/19/2021
Supplement Dates FDA Received07/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-