Model Number CLV-180
Device Problems Failure to Power Up (1476); Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was evaluated by olympus.The evaluation confirmed the reported event was due to an air pump flow rate which was out of specification.The device was unable to power up due to a faulty power supply and the evaluation uncovered that the high intensity mode was not working due to a worn-out scope socket slider switch.In addition, there was corrosion on the top cover and inside panel chassis and the non-olympus lamp meter was out of range at over 500+hours.The faulty parts were replaced and inspected to meet olympus functional standard.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
A user facility returned the olympus, evis exera ii xenon light source to olympus due to a report of insufficient and intermittent airflow from the device.Upon inspection and testing of the returned device, it was observed that the device has a power unit failure.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
The supplemental report is submitted to provide the result of the legal manufacturer¿s investigation.The device history record for the subject device was reviewed and it was verified that the device was manufactured in accordance with documented specifications.The legal manufacturer performed an investigation.The investigation presumed that the event occurred due to the age of the device.The air feed from the pump did not meet the standards because of the aging deterioration and malfunction of the air pump caused by repeated usage over a long period of time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11971291
MDR Text Key273572902
Report Number8010047-2021-07311
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170215544
UDI-Public04953170215544
Combination Product (y/n)N
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/13/2021
Initial Date FDA Received06/09/2021
Supplement Dates Manufacturer Received06/24/2021
Supplement Dates FDA Received07/01/2021
Is This a Reprocessed and Reused Single-Use Device? No