• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTAWIRE AND WIRECLIP TORQUER; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTAWIRE AND WIRECLIP TORQUER; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H802228240020
Device Problem Break (1069)
Patient Problem Foreign Body In Patient (2687)
Event Date 04/25/2021
Event Type  malfunction  
Event Description
Patient had significant vessel tortuosity.Using a 1.5mm burr for rotational artherectomy at 175,000 rpms the burr cut the rotafloppy wire.The fragment was unable to be retrieved and 3cm of retrained rotafloppy wire was stent up against the wall of the vessel (distal lad).Repeat angiogram did not show any obstruction at the site.Md believed patient anatomy contributed to the breakage of the wire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAWIRE AND WIRECLIP TORQUER
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key11982329
MDR Text Key255635834
Report Number11982329
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 06/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH802228240020
Device Catalogue NumberH802228240020
Device Lot Number26597428
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA06/04/2021
Date Report to Manufacturer06/11/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age22510 DA
-
-