• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P DISCOVERY ELBOW; DISC HUM 4X150MM LT FLANGED C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P DISCOVERY ELBOW; DISC HUM 4X150MM LT FLANGED C Back to Search Results
Model Number 540-34-150
Device Problems Loose or Intermittent Connection (1371); Appropriate Term/Code Not Available (3191)
Patient Problem Insufficient Information (4580)
Event Date 05/14/2021
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event Description
Second revision surgery - the humeral component was loose.Surgeon removed the humeral component from djo and put in a zimmer biomet segmental revision system humeral stem.The new condyle kit was to link that new stem with our ulnar stem.
 
Event Description
Revision surgery - the humeral component was loose.Surgeon removed the humeral component from djo and put in a zimmer biomet segmental revision system humeral stem.The new condyle kit was to link that new stem with our ulnar stem.
 
Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as loose stem.The previous surgery and the surgery detailed in this event occurred 10.7 months apart.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to loose stem.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC HUM 4X150MM LT FLANGED C
Manufacturer (Section D)
ENCORE MEDICAL L.P
9800 metric blvd
austin TX 78758 5445
MDR Report Key11984320
MDR Text Key255907615
Report Number1644408-2021-00566
Device Sequence Number1
Product Code JDC
UDI-Device Identifier00190446137733
UDI-Public(01)00190446137733
Combination Product (y/n)N
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 07/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number540-34-150
Device Catalogue Number540-34-150
Device Lot Number74058
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/14/2021
Initial Date FDA Received06/11/2021
Supplement Dates Manufacturer Received06/25/2021
Supplement Dates FDA Received07/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
540-00-000 LOT 84174
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age64 YR
-
-