THORATEC SWITZERLAND GMBH CENTRIMAG MOTOR, US; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
|
Back to Search Results |
|
Model Number 102956 |
Device Problems
Mechanical Problem (1384); Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/12/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
|
|
Event Description
|
It was reported that the perfusionist noticed that the flow probe was not reading.The flow probe was readjusted as well as unplugged and then re-plugged in with no resolution to the problem.The system was changed out successfully.Related manufacturer number of flow probe: 3003306248-2021-02970.Related manufacturer number of console: 3003306248-2021-02972.
|
|
Manufacturer Narrative
|
Manufacturer's investigation conclusion: the centrimag motor is being evaluated for the reported event of the flow probe not reading.The centrimag motor was not returned for analysis.Additional provided information communicated on (b)(6) 2021 stated that the issue was only with the flow probe.The serial number of the motor and console was unable to be provided upon request.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the device history records were unable to be reviewed due to the serial number of the motor being unknown.No further information was provided.The manufacturer is closing the file on this event.
|
|
Search Alerts/Recalls
|
|
|