Model Number N/A
Device Problem Inflation Problem (1310)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Expiration date - unknown due to unknown lot number.Udi - unknown due to unknown lot number.Implanted date: device was not implanted.Explanted date: device was not explanted.Device manufacture date - unknown due to unknown lot number.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.The production lot number was not provided by the user facility, which prevented a meaningful review of the device history record.
 
Event Description
The user facility reported that the tr band was releasing air post cardiac prior to adding the syringe.They had a tr band that was taken off a patient today.It was blown up to 12ml and held at a small amount of pressure assuming it would be the same as on a patient when its slowly deflating.After ten minutes they were only able to get 8mls of air out.The patient was in fine condition.The procedure outcome was a success.There was no patient injury, medical/surgical intervention required.
 
Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The complaint cannot be confirmed for air leakage since the sample was not returned for evaluation.Based on the information given, the exact root cause of the event cannot be determined.There is no indication that any manufacturing, design or quality system issues may have led to this event.Currently, no action is recommended since this risk evaluation is within the predetermined limits in the failure mode and effects analysis (fmea).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TR BAND
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
MDR Report Key11992847
MDR Text Key255991254
Report Number1118880-2021-00138
Device Sequence Number1
Product Code DXC
Combination Product (y/n)N
PMA/PMN Number
K152525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberTRB24-REG
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/21/2021
Initial Date FDA Received06/14/2021
Supplement Dates Manufacturer Received07/01/2021
Supplement Dates FDA Received07/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-