SMITH & NEPHEW, INC. LEGION CON ART INS 9MM SZ 3-4; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
Model Number 71420523 |
Device Problem
Incomplete or Inadequate Connection (4037)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/26/2021 |
Event Type
malfunction
|
Event Description
|
It was reported, that during a tka surgery, a legion con art ins 9mm size 3-4 did not lock into tibia bases tray.Surgery was completed, without delay, by using a back-up device.The patient was not harmed beyond what has been described.
|
|
Manufacturer Narrative
|
G3, h2,h3 and h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned insert confirms minor damage around the lock and base, more than likely from attempted insertion.A dimensional inspection was attempted but the device was too damaged from the attempted insertion to obtain accurate measurements.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|