BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH NORMAL SALINE SYRINGE; SALINE VASCULAR ACCESS FLUSH Back to Search Results
Model Number 306546
Device Problem Inability to Irrigate (1337)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/18/2021
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: it was reported by the consumer they were unable to push past 3ml.As a sample was unavailable for return, a thorough sample investigation could not be completed.It could be possible the customer had some product on the higher end of specification inducing more force than normal required to expel the solution.A device history record review was completed for provided material number 306546, lot number 1035833.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.Previous investigations have revealed that improper silicone application within the syringe barrel can create plunger resistance.Several quality initiatives have been implemented on our manufacturing line to ensure that the silicone application is properly applied during the manufacturing process.Additionally, a monitoring program is also in place to verify the silicone is uniformly applied to the syringe barrel.Based on the investigation and with no sample analysis the symptom reported by the customer could not be confirmed.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that the bd posiflush¿ normal saline syringe plunger would not push beyond the "3ml" mark.The following information was provided by the initial reporter: "the customer had several greater than 5 saline syringes that would not work.They aspirated and then attempted to inject the saline and were unable to push past 3ml.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH NORMAL SALINE SYRINGE
Type of Device
SALINE VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12007713
MDR Text Key256459673
Report Number1911916-2021-00559
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/31/2024
Device Model Number306546
Device Catalogue Number306546
Device Lot Number1035833
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/18/2021
Initial Date FDA Received06/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-