Model Number 0998-00-0800-53 |
Device Problem
Component Incompatible (1108)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/21/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp unit for the reported issue "unit failed drive manifold test", to fix the issue the fse replaced the transducer.The fse then performed a full pm with calibration, functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.(b)(6).
|
|
Event Description
|
It was reported that during a preventative maintenance (pm) service repair performed by a getinge field service engineer (fse), the cardiosave intra-aortic balloon pump (iabp) failed the drive manifold test.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Analysis of production: (3331) the device history record review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110) the overall 24 month product complaint trend data for the period jun-2019 through may-2021 was reviewed.There were no triggers identified for the review period.
|
|