• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 1.7 MM SQUARE DRIVER; INSTRUMENT, SPORTS MED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 1.7 MM SQUARE DRIVER; INSTRUMENT, SPORTS MED Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/11/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during a procedure the tip of the screwdriver broke off.No additional information is available.No adverse event reported.
 
Manufacturer Narrative
(b)(4) this follow-up report is being submitted to relay additional information.Updated: b4, b5, d9, g3, h2, h3, h4, h6 complaint sample was evaluated and the reported event was confirmed.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.Visual examination of the returned product identified the tip was fractured.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.7 MM SQUARE DRIVER
Type of Device
INSTRUMENT, SPORTS MED
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key12010579
MDR Text Key256549516
Report Number0001825034-2021-01805
Device Sequence Number1
Product Code HWR
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Type of Report Initial,Followup
Report Date 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number231200101
Device Lot NumberCA5C625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/11/2021
Initial Date FDA Received06/16/2021
Supplement Dates Manufacturer Received08/02/2021
Supplement Dates FDA Received08/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-