Device evaluated by mfr.: returned product consisted of an angiojet solent omni catheter.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually examined for damage or any irregularities.The catheter shaft showed multiple bends and kinks.There was severe tip damage and stretching.There was a broken hypotube located from the tip to 3cm proximal.Functional testing could not be completed due to the severe damage on the catheter shaft.No pressure reading could be reported.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|