MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 37601
Device Problems Electromagnetic Interference (1194); High impedance (1291); Failure to Interrogate (1332); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a manufacturer¿s representative (rep) who reported they had some difficulty taking initial impedances likely due to emi (being at the mri department), but they kept getting telemetry failure.During the call the rep went back into the application and into the impedance tab workflow where impedances were able to be completed.During the check high impedances were discovered.The impedance test was run at 3.0 v with the left lead being case/0 at 3,3339 ohms, but all other values were normal including the right lead.Due to this the patient was unable to move forward with having an mri.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key12016533
MDR Text Key266048007
Report Number3004209178-2021-09544
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00643169864191
UDI-Public00643169864191
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2020
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/16/2021
Initial Date FDA Received06/17/2021
Date Device Manufactured07/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
-
-