The patient was undergoing a thrombectomy procedure in the superficial femoral artery (sfa) using an indigo system aspiration catheter 8 (cat8), a non-penumbra sheath, and a guidewire.It should be noted that the patient had a steep iliac bifurcation.During the procedure, the physician completed one pass using the cat8 and the sheath.The physician then placed multiple stents from the popliteal artery to the sfa.While advancing the cat8 through the stents for the second pass, the cat8 became stuck on a stent strut.Subsequently, the physician torqued the cat8, and the cat8 broke into two pieces.Therefore, the physician removed part of the cat8 and the sheath.Next, the physician attempted to retrieve the broken distal end of the cat8 using two snare devices; however, the attempts were unsuccessful.The physician then attempted to use a balloon catheter to push the broken piece of the cat8 into the aorta to grab it; however, it was also unsuccessful.Therefore, an open surgery via cutdown technique was performed to remove the broken piece of the cat8.The procedure ended after the surgery.There was no report of an adverse effect to the patient.
|
Evaluation of the returned cat8 confirmed that the catheter was fractured.Based on the damage at the fracture, this break was likely due to the torquing of the device.The catheter being stuck on a stent may have contributed to the damage.Further evaluation revealed multiple kinks and ovalization on the fractured distal segment.This damage was incidental to the reported complaint and may have occurred during removal from the patient or post-procedure.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|