• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P7
Device Problem Problem with Sterilization (1596)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device in this report has not been returned to olympus medical systems corp.(omsc) for evaluation.After confirming the reported event, an asp (advanced sterilization products) representative provided the booster to the user facility and it was used properly at the user facility.The user facility began investigating whether there was an explanation from the asp representative or olympus representative, and the recognition of the staff in charge of reprocessing.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
The sales representative of olympus medical science sales corp.Confirmed that the user facility used the sterrad 100s sterilization system manufactured by johnson & johnson to sterilize the device without attaching the booster (ref15400) that needed to be connected to the device.The user facility may have been sterilized on a regular basis without the booster for at least two years.There was no report of patient injury and infection associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The exact cause of the reported event could not be conclusively determined, because the device has not been returned to olympus medical systems corp.(omsc).However, based on the information provided that the user facility used the sterrad 100s sterilization system manufactured by johnson & johnson to sterilize the device without attaching the booster (ref (b)(4)) that needed to be connected to the device, omsc determined that the reprocessing process by the user facility may have been inappropriate and insufficient.The instruction manual provides preventive measures against the reported failure mode.Device history record (dhr) review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12022631
MDR Text Key273307106
Report Number8010047-2021-07684
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P7
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/26/2021
Initial Date FDA Received06/18/2021
Supplement Dates Manufacturer Received06/28/2021
Supplement Dates FDA Received07/14/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2018
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-