Model Number 508-36-101
Device Problems Inadequacy of Device Shape and/or Size (1583); Unstable (1667); Appropriate Term/Code Not Available (3191)
Patient Problem Failure of Implant (1924)
Event Date 05/27/2021
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event Description
Revision surgery - patient's shoulder is unstable.Upsized to 40 sphere.
 
Manufacturer Narrative
The reason for this revision surgery was reported as instability.The previous surgery and the surgery detailed in this event occurred 6 months and 3 weeks apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this compliant was a revision surgery due to instability.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Type of Device
GLENOID, HEAD W/RETAINING SCREW, RSP, 36MM, NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin, tx 78758-5445
MDR Report Key12028692
MDR Text Key257203392
Report Number1644408-2021-00601
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912024761
UDI-Public(01)00888912024761
Combination Product (y/n)N
PMA/PMN Number
K051075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 07/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number508-36-101
Device Catalogue Number508-36-101
Device Lot Number869C3029
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/27/2021
Initial Date FDA Received06/18/2021
Supplement Dates Manufacturer Received06/18/2021
Supplement Dates FDA Received07/06/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
509-00-036 LOT 378P1309
Patient Outcome(s) Required Intervention;
Patient Age64 YR
-
-