• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL LP DISCOVERY ELBOW; DISC ULNA BEARING KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL LP DISCOVERY ELBOW; DISC ULNA BEARING KIT Back to Search Results
Catalog Number 114800
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Failure of Implant (1924)
Event Date 05/28/2021
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event Description
Revision surgery - surgeon determined pre-operatively that the ulna component was likely loose and scheduled revision surgery.The original condyle kit and ulna components were removed and the items were implanted.
 
Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was reported as loose component.The previous surgery and the surgery detailed in this event occurred 3.4 years apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to loose component.There were no findings during this evaluation that indicate that the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC ULNA BEARING KIT
Manufacturer (Section D)
ENCORE MEDICAL LP
9800 metric blvd
austin TX 78758 5445
MDR Report Key12028695
MDR Text Key257204622
Report Number1644408-2021-00595
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 07/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/11/2021
Device Catalogue Number114800
Device Lot Number459160
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/28/2021
Initial Date FDA Received06/18/2021
Supplement Dates Manufacturer Received07/20/2021
Supplement Dates FDA Received07/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
114700 LOT 201760; 114800 LOT 165870; 114990 LOT UNKNOWN
Patient Outcome(s) Other; Required Intervention;
Patient Age58 YR
-
-