• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US LUPINE BR DS W/ORTHCRD; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US LUPINE BR DS W/ORTHCRD; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 210712
Device Problem Device Damaged Prior to Use (2284)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/25/2021
Event Type  malfunction  
Manufacturer Narrative
Udi: b)(4).Investigation summary: according to the information provided, it was reported that during the arthroscopic right rotator cuff repair surgery, opened the packing (did not use), noted the anchor was deformed(as the attached photo shows).The complaint device is not being returned, therefore unavailable for a physical evaluation.However a photo was provided.Upon visual inspection of the photo, it could be observed that the anchor is bent, also the sutures are tensioning from the anchor to the handle's groove.The shaft has no structural anomalies.A manufacturing record evaluation was performed for the finished device 7l19094 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection of the photo, this complaint can be confirmed.The possible root cause can be attributed to the conditions of storage of the device.A combination of warmer temperature and tension can lead to a bent anchor.As per ifu 109002: store in a cool dry area.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Event Description
It was reported by the affiliate in (b)(6) that during an arthroscopic rotator cuff repair procedure of the right shoulder, it was observed that the lupine br ds w/orthcrd device was deformed upon opening its package.During in-house engineering evaluation, it was determined that the anchor was bent in the picture provided by the customer.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUPINE BR DS W/ORTHCRD
Type of Device
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key12031796
MDR Text Key257156033
Report Number1221934-2021-01928
Device Sequence Number1
Product Code MAI
UDI-Device Identifier10886705001101
UDI-Public10886705001101
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K150209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 06/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Model Number210712
Device Catalogue Number210712
Device Lot Number7L19094
Initial Date Manufacturer Received 06/18/2021
Initial Date FDA Received06/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-