Udi: b)(4).Investigation summary: according to the information provided, it was reported that during the arthroscopic right rotator cuff repair surgery, opened the packing (did not use), noted the anchor was deformed(as the attached photo shows).The complaint device is not being returned, therefore unavailable for a physical evaluation.However a photo was provided.Upon visual inspection of the photo, it could be observed that the anchor is bent, also the sutures are tensioning from the anchor to the handle's groove.The shaft has no structural anomalies.A manufacturing record evaluation was performed for the finished device 7l19094 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection of the photo, this complaint can be confirmed.The possible root cause can be attributed to the conditions of storage of the device.A combination of warmer temperature and tension can lead to a bent anchor.As per ifu 109002: store in a cool dry area.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
|
It was reported by the affiliate in (b)(6) that during an arthroscopic rotator cuff repair procedure of the right shoulder, it was observed that the lupine br ds w/orthcrd device was deformed upon opening its package.During in-house engineering evaluation, it was determined that the anchor was bent in the picture provided by the customer.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
|