Siemens has completed the investigation: the instrument was returned as an instrument only.No batteries or power supply were returned with the instrument.The d16 on the main board was the source of the burning smell but it is not a fire risk.D16 fails when incorrect power is connected to the device, either through plugging in the wrong power supply (device uses a common barrel connector) or using 3.5v rechargeable batteries (device specifies using common 1.5v aa batteries).In addition, this can be caused by any natural event causing a power surge.When opening the clinitek status+, faint smoke stains were noted on the plastics directly above component d16.In addition, failure of d16 was confirmed using a multimeter.
|