Model Number PT101 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Bradycardia (1751); Low Oxygen Saturation (2477)
|
Event Date 06/05/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of retrieving the subject pt101 airvo 2 humidifier for evaluation.We will provide a follow up report upon completion of our investigation.
|
|
Event Description
|
A healthcare facility in (b)(6) reported via a fisher & paykel healthcare (f&p) field representative that a patient desaturated to 23% spo2, experienced obtundation and was bradycardic at 15 bpm when the pt101 airvo 2 humidifier stopped providing flow.The patient was manually ventilated using a bag valve mask.The "patient responded quickly to manual ventilation and spo2 improved to 99%".No further patient consequence was reported.
|
|
Event Description
|
A healthcare facility in canada reported via a fisher & paykel healthcare (f&p) field representative that a patient desaturated to 23% spo2, experienced obtundation and was bradycardic at 15 bpm when the pt101 airvo 2 humidifier, set to 50lpm and 70% fio2, "stopped working and displayed message 'please turn off and restart'".The patient was manually ventilated using a bag valve mask and recovered to a stable condition at 99% spo2.There was no further patient consequence.
|
|
Manufacturer Narrative
|
(b)(4).Method: the complaint pt101 airvo 2 humidifier was returned to fisher & paykel healthcare (f&p) in new zealand where it was visually inspected, and performance tested.Results: visual inspection of the returned subject pt101 airvo 2 humidifier revealed no signs of external or impact damage.The error logs were reviewed showing that e192 error code was logged approximately at the time of the event.The unit was powered on, ran for four continuous hours and was found to be working correctly, with no error codes displayed.The subject device was opened for further investigation revealing no issues with the internal components of the unit.Conclusion: we are unable to determine what may have caused the disruption in therapy, as there was no fault found with the returned device.An e192 error code occurs when there is noise detected in the oxygen readings of the sensor.Flow continues to be delivered when e192 error code occurs.When it is triggered, an audible alarm will sound and a visual alarm will display "please turn off and restart".The airvo 2 humidifier user manual states: - "airvo 2 is for the treatment of spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases" - "the unit is not intended for life support." the user manual warns the user: - "appropriate patient monitoring must be used at all times.".
|
|
Search Alerts/Recalls
|