• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DBL OFFSET BROACH HDLE RT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DBL OFFSET BROACH HDLE RT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71360090
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2021
Event Type  malfunction  
Event Description
It was reported that the dbl offset broach hdle rt is not holding securely.As this happened in a non-surgical environment, there was not patient involvement.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DBL OFFSET BROACH HDLE RT
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12042927
MDR Text Key257489538
Report Number1020279-2021-05322
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010610362
UDI-Public03596010610362
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71360090
Device Catalogue Number71360090
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/07/2021
Initial Date FDA Received06/22/2021
Supplement Dates Manufacturer Received08/16/2021
Supplement Dates FDA Received08/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-