The reported complaint was confirmed.The srt noticed during oven testing that on the last cycle of four hours +/- 10% (24 minutes) at 30 degrees celsius +/- 5 degrees celsius, the venous probe failed the test with a temperature reading of 6.6 degrees celsius, which is out of specification.The srt replaced the venous probe due to intermittent thermistor readings during the oven testing and the unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|