• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II BCS XLPE ART ISRT SZ 1-2 RT 12MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II BCS XLPE ART ISRT SZ 1-2 RT 12MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74027214
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2021
Event Type  malfunction  
Event Description
It was reported that during a tka, the journey ii bcs xlpe art isrt size 1-2 right 12mm ((b)(4)) and could not be fully seated to journey tibia base np rt sz 2 ((b)(4)).There was a 15 minute delay and the procedure was finished using a smith & nephew back up.Patient was not harmed.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device does not confirm the stated failure mode.The returned insert has damage among the base, more than likely from attempted insertion.A dimensional inspection was attempted, but the device was too damaged from the attempted insertion to obtain accurate measurements.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS XLPE ART ISRT SZ 1-2 RT 12MM
Type of Device
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12046645
MDR Text Key258128074
Report Number1020279-2021-05364
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00885556178812
UDI-Public00885556178812
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74027214
Device Catalogue Number74027214
Device Lot Number19BM10529
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2021
Initial Date Manufacturer Received 06/09/2021
Initial Date FDA Received06/22/2021
Supplement Dates Manufacturer Received10/25/2021
Supplement Dates FDA Received10/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-