• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC STEALTHSTATION S8 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC STEALTHSTATION S8 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735669
Device Problems Display or Visual Feedback Problem (1184); Adverse Event Without Identified Device or Use Problem (2993); Data Problem (3196); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
Information references the main component of the system.Other relevant device(s) are: product id: 9735736, software version #: (b)(4).Software logs have been received, but not analyzed.Codes b21, c21, and d16 are applicable.No other products have been returned to medtronic for analysis.Codes b17, c20, and d15 are applicable.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system used during a functional endoscopic sinus surgery (fess).It was reported that after passing registration, the site received a 1.7mm metric, went to navigate, then returned and added a few trace patterns and then the model disappeared.They hit verify, then the screen went black and error message "system needs to restart." they rebooted system and the issue resolved.There was a reported delay to the procedure of less than one hour.There was no reported impact to the patient.
 
Manufacturer Narrative
H3: a medtronic representative went to the site to test the equipment.Testing revealed that the system performed as intended.The system then passed the system checkout and was found to be fully functional.B01, c19, d14 applicable codes h3: logs were received and analysis was found in sw- analysis determined it was confirmed that this was a known software anomaly.B01, c10, d02 applicable codes.Software version updated: 1.3.2 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key12058362
MDR Text Key258098781
Report Number1723170-2021-01671
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00763000360450
UDI-Public00763000360450
Combination Product (y/n)N
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735669
Device Catalogue Number9735669
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2021
Initial Date Manufacturer Received 06/21/2021
Initial Date FDA Received06/24/2021
Supplement Dates Manufacturer Received08/16/2021
Supplement Dates FDA Received08/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
"SEE H10...."; SEE H10; "SEE H10...."
Patient Age77 YR
Patient Weight91
-
-