Model Number 71692801
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/11/2021
Event Type  Injury  
Event Description
It was reported that, during a surgery, a sureshot targeter showed an error message when it was connected to the interface (the targeter was broken).Surgery was resumed, after a non-significant delay, with a back-up device (radiolucent drill).Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible probable cause could include but not limited to software not programmed correctly, software not updated, connection or mechanical issues.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESHOT TARGETER
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12063691
MDR Text Key260105949
Report Number1020279-2021-05440
Device Sequence Number1
Product Code OLO
UDI-Device Identifier03596010648167
UDI-Public03596010648167
Combination Product (y/n)N
PMA/PMN Number
K092497
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71692801
Device Catalogue Number71692801
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/11/2021
Initial Date FDA Received06/24/2021
Supplement Dates Manufacturer Received08/03/2021
Supplement Dates FDA Received08/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-