|
Model Number GF-558 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/28/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device was returned to the service center for evaluation; however, the device evaluation is still pending.This report will be supplemented accordingly following investigations.
|
|
Event Description
|
The user facility reported that prior to processing, the equipment was found broken.The user opened the package and noticed that the grasper was broken.The issue occurred prior to a stent removal procedure.The same device was not used to complete the procedure.No patient injury or harm was reported due to the event.No user injury reported.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The device was returned to olympus for evaluation.Upon evaluation of the device damage was able to be confirmed.The inner sheath was kinked and prevented the grasper from being fully closed.The device was returned to the oem (original equipment manufacturer) for further evaluation.The oem confirmed the damage and the damage is consistent with device reuse and reprocessing.The true lot number was not provided, the date code on the device is eb.This device was manufactured in may 2018, which is lot number 40246 or 40247.The oem performed a dhr (device history record) review on the two possible affected lots and the inspection results met the acceptance criteria and no unusual results or defects were noted.Olympus will continue to monitor the field performance of this device.
|
|
Search Alerts/Recalls
|
|
|