MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 32MM X 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 32MM X 50MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71337650
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/12/2021
Event Type  malfunction  
Event Description
It was reported that, during thr surgery, an r3 20 deg xlpe acet lnr 32mm x 50mm couldn't be locked after trying several times.The procedure was completed using a s+n back-up device.The event resulted in an unplanned increase of 10 minutes of surgery time, increased the patient's blood loss and risk of infection.Patient was not harmed.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The device has signs of damage from the attempted insertion.The returned part was evaluated for critical to quality features.All measured features are within tolerance per associated drawing print.At this point, the cause of the complaint cannot be contributed to the dimensions of the returned product.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 32MM X 50MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12077050
MDR Text Key258694305
Report Number1020279-2021-05477
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010598486
UDI-Public03596010598486
Combination Product (y/n)N
PMA/PMN Number
K113848
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71337650
Device Catalogue Number71337650
Device Lot Number19MM04447
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/12/2021
Initial Date FDA Received06/28/2021
Supplement Dates Manufacturer Received09/17/2021
Supplement Dates FDA Received09/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-