• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL INC. MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE; ABSORBENT, CARBON DIOXIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL INC. MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE; ABSORBENT, CARBON DIOXIDE Back to Search Results
Model Number MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2021
Event Type  malfunction  
Manufacturer Narrative
At this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported that the medisorb¿ multi-absorber original, disposable experienced broken cartridges.The issue occurred during patient-use and the customer confirmed that there was no patient harm associated with the reported event.
 
Manufacturer Narrative
Result of investigation: sample received for evaluation.Sample was leak tested and although it was found to be leaking at 10.9ml/min over 100 cmh2o, this was within vyaire's specification of < 20ml/min in accordance with pqas-8003138 et al-ca.Therefore no root cause could be determined.All units packed passed 100% leak test and visual inspection.A review of the batch documentation for lot l01a-00256 found no issues that could be linked with damage failures on this batch.The pallet and carton for canisters clearly states that they are fragile and to handle with care.Devices are packaged in a carton that meets the vyaire drop testing specification.Capa 21-013 has been opened up due to leak failures and these type of complaints are being tracked on our complaints system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE
Type of Device
ABSORBENT, CARBON DIOXIDE
Manufacturer (Section D)
VYAIRE MEDICAL INC.
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
MOLECULAR PRODUCTS LTD
parkway harlow business park
harlow, essex, CM19 5FR
UK   CM19 5FR
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
MDR Report Key12081301
MDR Text Key263511896
Report Number3002807637-2021-00019
Device Sequence Number1
Product Code CBL
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMEDISORB¿ MULTI-ABSORBER ORIGINAL, DISPOSABLE
Device Catalogue Number8003138
Device Lot NumberL01-00256
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/02/2021
Initial Date FDA Received06/28/2021
Supplement Dates Manufacturer Received11/21/2021
Supplement Dates FDA Received12/14/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-