U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SINGAPORE OPERATIONS ENSURA DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SINGAPORE OPERATIONS ENSURA DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number EN1DR01
Device Problems Mechanical Problem (1384); Device Contamination with Body Fluid (2317)
Patient Problem Hematoma (1884)
Event Date 06/15/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that two days post implant, the patient experienced a hematoma in the implantable pulse generator (ipg) pocket.On revision, an absence of insulation was noted on the setscrew.The ipg was explanted and replaced. no further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSURA DR MRI SURESCAN
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SN  486056
MDR Report Key12082548
MDR Text Key258871420
Report Number3008973940-2021-02212
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/14/2022
Device Model NumberEN1DR01
Device Catalogue NumberEN1DR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2021
Initial Date Manufacturer Received 06/25/2021
Initial Date FDA Received06/29/2021
Supplement Dates Manufacturer Received08/04/2021
Supplement Dates FDA Received08/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age65 YR
-
-