• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. DRIVER IMPACTOR-FEMORAL; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. DRIVER IMPACTOR-FEMORAL; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/08/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that an impactor fractured while impacting the implant.It fractured after the implant was already impacted and it did not delay the case.Attempts to obtain additional information have been made; however, no more information is available at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: b4, b5, g3, g6, h1, h2, h3, h6, and h10.Corrected: h4 visual examination of the returned product identified signs of use and it is fractured.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.The root cause of the reported issue is attributed to wear and tear from repeated use of the instrument for twenty years.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVER IMPACTOR-FEMORAL
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key12085281
MDR Text Key259017702
Report Number0001822565-2021-01772
Device Sequence Number1
Product Code HWR
UDI-Device Identifier00889024186262
UDI-Public(01)00889024186262(10)68770800
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 07/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00133602800
Device Lot Number68770800
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/08/2021
Initial Date FDA Received06/29/2021
Supplement Dates Manufacturer Received07/23/2021
Supplement Dates FDA Received07/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age50 YR
-
-