• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII MIS DCF ALIGN GDE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII MIS DCF ALIGN GDE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71441144
Device Problems Break (1069); Defective Device (2588); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2021
Event Type  malfunction  
Event Description
It was reported that, a gii mis dcf align gde is damaged.As this was noticed during inspection activities, no patient was involved.During the investigation pieces were found broken off and the device shows signs of extensive wear.
 
Manufacturer Narrative
H3, h6: the device, intended for use used in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be fractured and the fractured pieces were not returned, rendering the device inoperable.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII MIS DCF ALIGN GDE
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12087767
MDR Text Key259203012
Report Number1020279-2021-05546
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010497161
UDI-Public03596010497161
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71441144
Device Catalogue Number71441144
Device Lot Number07GM16932
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2021
Initial Date Manufacturer Received 06/15/2021
Initial Date FDA Received06/29/2021
Supplement Dates Manufacturer Received08/17/2021
Supplement Dates FDA Received08/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-