Model Number DTPB2QQ
Device Problems Mechanical Problem (1384); Misconnection (1399)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient presented after an alert tone due to short ventricular intervals and high undefined pacing impedance on the right ventricular lead approximately two weeks post-implant.Noise was also identified upon further investigation.Reprogramming was performed and lead replacement was planned.During the procedure, the physician was able to replicate the noise while the lead was still in the cardiac resynchronization therapy defibrillator (crt-d).The lead was disconnected and the physician noticed that the lead came out with ease even though it appeared to be pushed all the way through the crt-d header.The lead was connected to an analyzer but noise could not be reproduced and the lead was functioning nominally.The device was replaced due to a suspected issue with the right ventricular port or grommet and setscrew.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBALT HF QUAD CRT-D MRI SURESCAN
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ  1131
MDR Report Key12091079
MDR Text Key259129147
Report Number9614453-2021-02664
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00763000178178
UDI-Public00763000178178
Combination Product (y/n)N
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/14/2022
Device Model NumberDTPB2QQ
Device Catalogue NumberDTPB2QQ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2021
Initial Date Manufacturer Received 06/10/2021
Initial Date FDA Received06/30/2021
Supplement Dates Manufacturer Received08/05/2021
Supplement Dates FDA Received08/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
6935M62 LEAD, 459888 LEAD, 5076-52 LEAD; 6935M62 LEAD, 459888 LEAD, 5076-52 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age73 YR
-
-