MAUDE Adverse Event Report: AESCULAP INC PRESTIGE ATRA GRASPER DBL-ACT 5MM; REUSABLE INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP INC PRESTIGE ATRA GRASPER DBL-ACT 5MM; REUSABLE INSTRUMENTS Back to Search Results
Model Number 8360-10
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/14/2019
Event Type  malfunction  
Manufacturer Narrative
During a comprehensive review of the reported event, it was noted that the failure mode was described as the device not being able to close all the way.As this failure mode was escalated to a reportable malfunction in april 2018, the initial information and subsequent investigation results require mdr reporting.A supplier corrective action request (scar) was initiated due to an adverse trend observed for devices exhibiting failure at the thumb-loop assembly joint.The supplier evaluated the malfunction by looking at devices with a lot number beginning with "m." as a result of their findings, the push rod fixture was redesigned to increase the clearance, which ensured that the fixture appropriately stressed the entirety of the soldering joint.Upon implementing the new fixture, the supplier tested returned non-conforming samples, and verified that the soldering no longer hung up on the new fixture, as had been observed on the previous one.The complaint device was returned to the manufacturer for physical evaluation.A visual examination was performed which confirmed separation at the proximal weld.All returned non-conforming devices exhibited the same weld failure; a clean break occurred from the solder to the rotator housing.An investigation of the device manufacturing records was conducted by the manufacturer for the lot # of the device in question.No non-conformances were reported.All device history records (dhr) are reviewed and released according to documented procedures and a device is not released if it does not meet requirements or is nonconforming.Additionally, historical scrap rates were reviewed with no increase observed in scrap related to the complaint issue.In addition to the redesign of the soldering fixture, the supplier reviewed the work instruction for the torch soldering operation and identified improvement opportunities.The original work instruction was used to better define the soldering process with a more focused emphasis on the following: equipment startup and shutoff operations, clear imagery of acceptable soldered subassemblies, and clarified cleaning operations for components prior to soldering.A second dedicated work instruction was implemented to better define the attribute inspection criteria for the brazing process used for the solder between the thumb loop and rotator block.The investigation into the cause of the reported problem was able to confirm the failure mode of a proximal weld failure.This event likely occurred due to inadequacies in the defined production process which limited the device performance.Therefore, the most probable root cause is considered to be manufacturing related.In addition to a supplier corrective action request (scar) being initiated, a corrective action/preventive action (capa) was opened by aesculap inc.For further evaluation of the design transfer of this device.
 
Event Description
It was reported that there was an issue with a prestige grasper.The operating room (or) staff was setting up for a case and noticed that the instrument was not functioning properly.The wave-grasper did not close all the way.The device was sent to sterile processing and forwarded to repair.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESTIGE ATRA GRASPER DBL-ACT 5MM
Type of Device
REUSABLE INSTRUMENTS
Manufacturer (Section D)
AESCULAP INC
3773 corporate parkway
center valley PA 18034
Manufacturer (Section G)
AESCULAP INC
3773 corporate parkway
center valley PA 18034
Manufacturer Contact
lindsay chromiak
3773 corporate parkway
center valley, PA 18034
MDR Report Key12094037
MDR Text Key261267191
Report Number2916714-2021-00133
Device Sequence Number1
Product Code NWV
UDI-Device Identifier04046955083374
UDI-Public4046955083374
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 06/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8360-10
Device Catalogue Number8360-10
Device Lot NumberM45041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/14/2019
Initial Date FDA Received06/30/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-