• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M Back to Search Results
Model Number BI70000028120
Device Problem Temperature Problem (3022)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2021
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the generator batteries were weak.The batteries were replaced, the imaging system then passed the system checkout and was found to be fully functional.Other relevant device(s) are: product id: bi71000125, serial/lot #: unk ; product id: bi71000126, serial/lot #: unk.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging device being used outside of a procedure.It was reported that site called in reporting that the site is experiencing a thermal event via odor from the imaging system.There was no patient present when this issue occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC (LITTLETON)
300 foster st
littleton MA 01460
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key12100638
MDR Text Key259418082
Report Number3004785967-2021-00862
Device Sequence Number1
Product Code OXO
UDI-Device Identifier00643169353428
UDI-Public00643169353428
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092564
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70000028120
Device Catalogue NumberBI70000028120
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/10/2021
Initial Date FDA Received07/01/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-