MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 955467
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2021
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment with a prismaflex m150, an access extremely positive alarm was triggered due to a deformed luer connector.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the actual device was not available; however, photos were provided for evaluation.Visual inspection observed that the return male luer lock was broken.The reported condition was verified.The cause was design related.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M150
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key12104001
MDR Text Key259601440
Report Number8010182-2021-00212
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414123451
UDI-Public(01)07332414123451(10)
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2022
Device Catalogue Number955467
Device Lot Number20L0106
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/09/2021
Initial Date FDA Received07/01/2021
Supplement Dates Manufacturer Received08/30/2021
Supplement Dates FDA Received09/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NI.
-
-