Model Number MAJ-891
Device Problems Break (1069); Degraded (1153)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to omsc for evaluation but was returned to olympus europa (b)(4).Olympus europa (b)(4) checked the subject device and found the reported phenomena.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at olympus europa (b)(4), it was found that the lever with the stopper was detached from the subject device, and there was corrosion on the tightening ring.The date of event is unknown.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the subject device evaluation result.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation but was returned to olympus europa se & co.Kg (oekg).Omsc could not review the manufacturing history record (dhr) because the lot number was unknown, but normally, products that meet the shipping standards are shipped from the factory.Based upon the information from oekg, omsc considered that the reported phenomena were attributed to the followings.- the lever with the stopper was detached from the subject device; a) the strong chemical stress was applied to the subject device.B) the strong physical stress was applied due to such as forcibly attaching and detaching parts.The deterioration of the glue due to the combination chemical stress during reprocessing such as a) and b) and the mechanical stress during assembly.- there was corrosion on the tightening ring; the occurrence of crevice corrosion, the glue that fixed the tightening ring was corroded and discolored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCEPS/IRRIGATION PLUG (ISOLATED TYPE)
Type of Device
FORCEPS/ IRRIGATION PLUG
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12116248
MDR Text Key270049973
Report Number8010047-2021-08423
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170063114
UDI-Public04953170063114
Combination Product (y/n)N
PMA/PMN Number
K912120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberMAJ-891
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/10/2021
Initial Date FDA Received07/05/2021
Supplement Dates Manufacturer Received08/17/2021
Supplement Dates FDA Received09/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-