Model Number 71440005
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2021
Event Type  malfunction  
Event Description
It was reported that, during a field inspection, a screw fell off a femoral implant impactor.As this was noticed in a non-surgical environment, there was not any patient involved.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The device is fractured and the fractured piece was returned, rendering the device inoperable.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL IMPLANT IMPACTOR
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12119184
MDR Text Key260102099
Report Number1020279-2021-05718
Device Sequence Number1
Product Code JDG
UDI-Device Identifier00885556034156
UDI-Public00885556034156
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440005
Device Catalogue Number71440005
Device Lot Number15BM12796
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2021
Initial Date Manufacturer Received 06/22/2021
Initial Date FDA Received07/05/2021
Supplement Dates Manufacturer Received09/13/2021
Supplement Dates FDA Received09/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-