MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
|
Back to Search Results |
|
Model Number BI70000028230 |
Device Problems
Failure to Run on Battery (1466); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/15/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved when the reported issue occurred.A medtronic representative went to the site to test the equipment.Testing revealed that motion batteries for the imaging system were replaced.The imaging system then passed the system checkout and was found to be fully functional.Concomitant medical products: other relevant device(s) are: product id: b i71000125, serial/lot #: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used during a sacroiliac and thoracolumbar procedure.It was reported that the imaging system could not perform image acquisitions.It was reported that when the rotor moved when attempting a 3d image acquisition, but the scan could not be executed.The surgeon opted to complete the procedure without the use of the imaging system and medtronic navigation system.There was a reported delay to the procedure of more than 1 hour due to this issue.There was no reported impact on patient outcome.Troubleshooting found that the event logs indicated an issue with the battery charge.
|
|