Model Number 500AHCT
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The srt replaced the aux board.The unit operated to the manufacturer's specifications.
 
Event Description
The service repair technician (srt) reported that during testing of the device at the service center, the blood parameter monitor (bpm) had heat damage on the c130 capacitor on the auxiliary (aux) printed circuit board assembly (pcba) board.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 500
Type of Device
MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key12127863
MDR Text Key260242101
Report Number1828100-2021-00231
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
PMA/PMN Number
K133658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number500AHCT
Device Catalogue Number500AHCT
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/14/2021
Initial Date FDA Received07/07/2021
Supplement Dates Manufacturer Received07/23/2021
Supplement Dates FDA Received08/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-