• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-AW
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2021
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) was informed that there was water sample problem and positive culture results for rinse water of oer-aw.There was no report of patient infection associated with the event.In addition, omsc received the following information.- the user replaced all filters of oer-aw, and performed dis line after testing lcg with test strip.- after that drain lcg, connected to sampling tube and performed program 2 reprocessing.- finally function selected rinse to get the sampling.- but the result still found mold.
 
Manufacturer Narrative
The subject device was returned to local service department of olympus.The exact cause has been under investigation.If additional information is received, this report will be supplemented.
 
Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation.Therefore, the exact cause of the reported event could not be conclusively determined.The manufacturing record was reviewed and found no irregularities.Omsc presumed the following possible causes.Continued to use the water filter for longer than the period recommended by olympus.Disinfection of water supply tube was not carried out.The disinfectant was deteriorated.Contamination during sampling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12133231
MDR Text Key280753469
Report Number8010047-2021-08560
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 10/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberOER-AW
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/16/2021
Initial Date FDA Received07/08/2021
Supplement Dates Manufacturer Received09/17/2021
Supplement Dates FDA Received10/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-