Catalog Number J2C1702 |
Device Problem
Backflow (1064)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/15/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a small volume infusor had backflow from the filling port.This issue was discovered during filling with unspecified medication.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.A visual inspection was performed and two vertical cracks were observed located on the fill port.The sample was leak tested with green color water, and a leak was observed coming out of the two vertical cracks.No evidence of "backflow" was observed from the inside of the fill port.The cause of the two cracks could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|