Model Number XS PST |
Device Problem
Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Occupation is lay user/patient.The meter was requested for investigation.
|
|
Event Description
|
The initial reporter complained of a display issue with coaguchek xs meter serial number (b)(4).The customer stated segments were missing from the results field of the display.During a display check the meter displayed 000 instead of 888.There was no allegation that any test results had been misinterpreted.
|
|
Manufacturer Narrative
|
The reporter's meter was returned for investigation.Upon investigation of the returned meter, a defect in the circuit board or an electronic component of the circuit board was found.Medwatch fields d9 and h3 have been updated.
|
|