• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STONE CONE; DISLODGER, STONE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STONE CONE; DISLODGER, STONE, FLEXIBLE Back to Search Results
Model Number M0063903200
Device Problems Difficult to Open or Close (2921); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/13/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The returned stone cone was analyzed, and a visual inspection note that the blue outer sheath was detached from the working length and coil.Additionally, there was a tear and signs of damage at the distal end of the blue sheath.The observed condition would not be able to recapture the coil.No other problems with the device were noted.The reported event was confirmed.Based on all available information, the investigation concluded that it is likely that the difficulty experienced by the user occurred while testing the device prior to use.The damage is consistent with excessive force applied while attempting to sheath the coil.Therefore, the most probable root cause is unintended use error.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event Description
It was reported to boston scientific corporation that a stone cone nitinol retrieval coil was used in the middle and upper ureter during a ureteral lithotomy under rigid ureteroscopy procedure performed on (b)(6) 2021.During the procedure, before the device could be used, the coil of the stone cone could not close again outside of the patient.Another stone cone device was used to complete the procedure.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.The investigation results revealed that the sheath was torn at the distal section; therefore, this is now an mdr reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STONE CONE
Type of Device
DISLODGER, STONE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
LAKE REGION MEDICAL
31c butterfield trail
el paso TX 79906
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key12150087
MDR Text Key261474136
Report Number3005099803-2021-03434
Device Sequence Number1
Product Code FGO
UDI-Device Identifier08714729430223
UDI-Public08714729430223
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/15/2023
Device Model NumberM0063903200
Device Catalogue Number390-320
Device Lot Number0005225785
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/16/2021
Initial Date FDA Received07/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
Patient Weight79
-
-