Model Number 130790030 |
Device Problems
Off-Label Use (1494); Osseointegration Problem (3003)
|
Patient Problems
Inadequate Osseointegration (2646); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 06/22/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
The patient was revised to address infection and loosening of the metaglene at the bone to implant interface.The bone and underside had come loose.Doi: (b)(6) 2020, dor: (b)(6) 2021, left shoulder.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5 and h6 (device) if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
No instrumentation broke.Surgeon purposely used a right in a left shoulder due o unique anatomy.It was assuming loosening was caused by infection.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|